medical-device-software-compliance

Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure, 510(k)/De Novo/PMA pathway selection, SOUP/OTS assessment, verification and validation planning, cybersecurity for medical devices, PCCP for AI/ML devices, design reviews, traceability matrices, or post-market surveillance. Triggers include "SaMD", "medical device software", "IEC 62304", "ISO 14971", "ISO 13485", "21 CFR 820", "QMSR", "510(k)", "De Novo", "PMA", "design history file", "DHF", "software safety classification", "SOUP list", "risk management", "hazard analysis", "design review", "V&V protocol", "FDA submission", "EU MDR", "clinical evaluation", "PCCP", "predetermined change control", "GMLP", "cybersecurity premarket", "Part 11", "design controls", "design inputs", "design outputs", "traceability matri

Installs
89
Last 24h
—
Trending rank
—
Security
Not audited

Install

Works with every agent the skills CLI supports.

npx skills add https://github.com/awslabs/hcls-agent-skills --skill medical-device-software-compliance
  1. 01

    Install

    Run the command in your project. The CLI asks which agents to install for.

  2. 02

    Describe the task

    Ask your agent for something this skill covers. It loads medical-device-software-compliance automatically when relevant.

  3. 03

    Review the result

    The agent follows the SKILL.md steps, scripts and references bundled with the skill.

SKILL.md

The instructions your agent reads when the skill is triggered.

SaMD Compliance — Reasoning Skill

Overview

You are an expert in Software as a Medical Device (SaMD) regulatory compliance. When the user asks about FDA software device submissions, IEC 62304 lifecycle, ISO 14971 risk management, design controls, or global regulatory strategy, apply the decision frameworks below.

Usage

  • Invoke when evaluating SaMD regulatory compliance for FDA, EU MDR, or multi-market submissions
  • Use when planning design controls, risk management, V&V, or submission strategy
  • Activate for safety classification, SOUP assessment, or AI/ML device lifecycle questions

Response Format

  • Lead with the direct recommendation or classification (≤3 sentences)
  • Structure as: recommendation → justification (citing specific standards/clauses) → caveats
  • Use tables for comparisons; bullet points for criteria lists
  • Omit background the user already knows — they asked the question
  • Target: 200-400 words unless the user requests exhaustive detail

The decision trees and frameworks in this skill are for internal reasoning only. Apply them to reach your conclusion, but do not reproduce them in your response. Present only the final recommendation with supporting evidence.


1. IEC 62304 Safety Classification Decision Tree

Does the software system contribute to a hazardous situation?
├─ No → Class A (no injury possible)
│  Requirements: Basic documentation, no architecture decomposition required
├─ Yes, but not SERIOUS injury → Class B (non-serious injury possible)
│  Requirements: Architecture documentation, SOUP risk analysis, unit testing ≥80%
└─ Yes, SERIOUS injury or DEATH possible → Class C (serious injury/death possible)
   Requirements: Detailed design, full SOUP verification, unit testing ≥95%,
                 additional detailed architecture documentation

Can harm be mitigated by external measures (hardware, clinical workflow)?
├─ Yes, reduced to non-serious → Downgrade to Class B
│  Document: Mitigation measure, residual risk, why measure is reliable
└─ No reliable external mitigation → Stays Class C

Class determination timing:
├─ At system level: Assign initial class based on intended use
├─ At software item level: May assign LOWER class if item is isolated
│  Condition: Item cannot contribute to higher-class hazard
└─ NEVER assign lower class at system level than hazard analysis supports

2. FDA Regulatory Pathway Decision Tree

Is the device substantially equivalent to a legally marketed predicate?
├─ Yes, same intended use + same/different technology (equiv. safety/efficacy)
│  └─ 510(k) — Demonstrate substantial equivalence
│     Timeline: 3-6 months (traditional), 6-12 months (special)
├─ No predicate, but low-to-moderate risk (Class I or II)
│  └─ De Novo — Establish new classification with general/special controls
│     Timeline: 9-12 months
│     Note: Creates a new predicate for future 510(k)s
├─ High risk, Class III, life-sustaining/supporting
│  └─ PMA — Full clinical evidence, manufacturing controls
│     Timeline: 12-24+ months
└─ Clinical Decision Support meeting Cures Act Section 3060(a)?
   All 4 criteria met:
   1. Not intended to acquire/process/analyze medical images/signals
   2. Intended for healthcare professionals (not patients)
   3. Intended to enable HCP to independently review basis
   4. Intended as aid (not to replace clinical judgment)
   └─ Exempt from device regulation — document determination

3. Design Controls Lifecycle (21 CFR 820.30 / QMSR)

PhaseKey OutputIEC 62304 MappingBlocking Gate
User NeedsUser Needs Document§5.2 Software requirements processNone
Design InputSRS (Software Requirements Spec)§5.2.1-5.2.6 RequirementsDR1
Design OutputArchitecture + Detailed Design§5.3 Software architectural designDR2
VerificationTest protocols + reports§5.5 Software integration testing, §5.6 System testingDR3
ValidationClinical validation evidence§5.7 Software releaseSubmission
TransferManufacturing/deployment procedures§5.8 Maintenance processRelease
ChangesDesign change assessment§5.2 (re-entry)Per-change

Critical rule: Design Reviews (DR1, DR2, DR3) are BLOCKING for Class B/C devices. Each requires documented attendees, findings, and formal sign-off.

4. SOUP/OTS Assessment Framework

Risk-Based SOUP Classification

Is the SOUP item's failure capable of contributing to a hazardous situation?
├─ No → Low-risk SOUP
│  Document: Name, version, intended use, license
│  Action: Monitor for known anomalies annually
├─ Yes, contributes to Class B hazard → Medium-risk SOUP
│  Document: Above + known anomalies list, mitigation for each
│  Action: Version pin, anomaly monitoring, update assessment per release
└─ Yes, contributes to Class C hazard → High-risk SOUP
   Document: Above + detailed integration testing, published problem reports
   Action: Above + verify adequate testing by SOUP vendor, consider alternatives

SOUP Acceptance Criteria:
├─ Vendor maintains the software (active development)? → Preferred
├─ Known anomaly list published? → Required for Class B/C
├─ CVE history acceptable? → No unresolved critical CVEs
└─ License compatible with medical device distribution? → Required

SOUP vs Custom Decision

FactorUse SOUPBuild Custom
Well-validated open source library✅
Critical safety function with no validated library✅
Standard utility (logging, HTTP, encryption)✅
Regulatory precedent for the library✅
SOUP vendor unresponsive to anomaly reports✅
Time-to-market critical, library mature✅

5. ISO 14971 Risk Management Process

Severity Classification

LevelDescriptionExamplesAcceptable Probability
S1NegligibleInconvenience, no injuryAny
S2MinorTemporary minor injuryOccasional
S3SeriousInjury requiring interventionRemote
S4CriticalPermanent impairmentImprobable
S5CatastrophicDeathImprobable

Risk Control Priority (ISO 14971 §7.1)

Risk exceeds acceptability threshold?
├─ Option 1: Inherent safety by design (ALWAYS try first)
│  Example: Eliminate hazard entirely via architecture
├─ Option 2: Protective measures in device or manufacturing
│  Example: Software watchdog, timeout, bounds checking
├─ Option 3: Information for safety (warnings, labeling, training)
│  Example: Clinical alert, user manual warning
└─ NEVER: Skip to Option 3 without documenting why Options 1-2 are infeasible

Residual Risk Assessment:
├─ Individual residual risk acceptable? → Document and proceed
├─ Individual unacceptable but reducible? → Apply additional controls
└─ Overall residual risk vs. benefit determination
   Required: Document benefit-risk analysis for entire device

Hazard Analysis Common Pitfalls

  1. Wrong: Listing software bugs as hazards Right: Trace bug → hazardous situation → harm. The HARM is what matters.
  2. Wrong: Copying generic hazard lists without device-specific analysis Right: Analyze YOUR device's specific failure modes in YOUR clinical context
  3. Wrong: Assuming software cannot cause physical harm Right: Delayed diagnosis, wrong treatment recommendation, alert fatigue ALL cause harm
  4. Wrong: Risk controls that rely entirely on user vigilance Right: Design-level controls first; training/warnings are last resort

6. V&V Planning by Safety Class

ActivityClass AClass BClass C
Requirements traceability✅✅✅
Unit testingOptional≥80% coverage≥95% coverage
Integration testingOptional✅✅ + fault injection
System testing✅✅✅ + stress/boundary
Regression testingOn changeFull suiteFull suite + risk-based
Clinical validationIf claimedIf claimed✅ Mandatory
Cybersecurity testingIf connected✅ Penetration test✅ Full threat model
Usability validationIf HCP user✅ Formative✅ Summative

V&V Failure Recovery Decision Tree

Test fails acceptance criteria
├─ Safety-related requirement?
│  ├─ Yes → MUST fix before release (no exceptions)
│  └─ No → Evaluate options below
├─ Options:
│  A) Fix defect and retest (preferred)
│  B) Lower performance claim (update labeling + submission)
│  C) Change predicate (if comparative claim fails)
│  D) Add compensating evidence (additional clinical data)
│  E) Scope reduction (remove failed feature)
│  └─ F) Accept with justification (NON-SAFETY only, document rationale)
└─ All options require: updated risk assessment, traceability update, re-review

7. Cybersecurity Requirements (FDA 2023 Guidance)

Threat Modeling Minimum Scope

CategoryMust AddressExample Threats
ConfidentialityPHI exposureUnencrypted API, debug endpoints
IntegrityData/algorithm tamperingModel poisoning, input manipulation
AvailabilityDenial of serviceResource exhaustion, dependency failure
AuthenticationUnauthorized accessDefault credentials, session hijacking
Update mechanismSecure patchingUnsigned updates, rollback attacks

SBOM Requirements (per FDA Refuse-to-Accept checklist)

  • Format: CycloneDX 1.5+ or SPDX 2.3+
  • Must include: ALL direct + transitive dependencies
  • Per component: name, version, supplier, license, known CVEs
  • Update frequency: Every release + within 24h of critical CVE disclosure
  • FDA will RTA (Refuse to Accept) submissions without SBOM as of Oct 2023

8. AI/ML SaMD Lifecycle (PCCP Framework)

FDA Total Product Lifecycle for AI/ML

Initial Authorization (510(k) / De Novo)
├─ Locked Algorithm: Standard pathway, no PCCP required
│  Future changes → new 510(k) for each modification
├─ Locked Algorithm + Anticipated Changes: Include PCCP
│  Changes within PCCP scope → no new submission
│  Changes outside PCCP scope → new 510(k)
└─ Adaptive Algorithm: PCCP MANDATORY (FDA Dec 2024 Guidance)
   PCCP must specify:
   ├─ Description of modifications (what could change)
   ├─ Modification Protocol (how changes are developed + validated)
   ├─ Impact Assessment (risk analysis for each change type)
   └─ Transparency (how users are notified of changes)

Good Machine Learning Practice (GMLP) — 10 Principles

#PrincipleVerification Method
1Multi-disciplinary teamDocument team roles (clinical + engineering + regulatory)
2Good Software Engineering PracticeIEC 62304 compliance
3Representative clinical dataDataset demographics vs. intended population
4Independent training/test/validation setsNo data leakage verification
5Reference datasets (ground truth)Adjudication process documented
6Model design fits intended useArchitecture justification document
7Clinically relevant performance metricsSensitivity/Specificity/PPV/NPV/AUC
8Testing across patient subgroupsBias analysis across demographics
9Clear user informationLabeling re: capabilities + limitations
10Deployed model monitoringDrift detection + performance tracking

Performance Monitoring Thresholds

MetricAction ThresholdEscalation
Overall performance (AUC/F1)>5% degradation from validationInvestigate root cause
Subgroup performance>10% gap vs. majority subgroupBias review
False negative rate (safety)Any increase >2%Immediate CAPA
Data distribution shiftOOD rate >15% of inputsRetraining evaluation
Alert volume (CDS devices)>20 alerts/clinician/shiftAlert fatigue review

9. Traceability Requirements

Bidirectional Traceability Matrix Structure

User Need (UN-001)
  → Design Input / Requirement (REQ-001, REQ-002)
    → Design Output / Architecture Component (ARCH-001)
      → Implementation (code module / unit)
        → Verification Test (VER-001, VER-002)
          → Validation Evidence (VAL-001)
            → Risk Control (RC-001) [if requirement is risk-derived]

Traceability Completeness Rules

CheckBlocking?Description
Orphan requirement (no parent UN)YesEvery REQ must trace to a user need
Unverified requirementYesEvery REQ must have ≥1 verification test
Unimplemented risk controlYesEvery RC in hazard analysis must trace to code
Dead code (no tracing requirement)WarningMay indicate scope creep or incomplete specs
Missing validation linkClass B/CEach user need must trace through to validation

10. Design Review Checklist (DR1 / DR2 / DR3)

DR1 — Inception Exit (Before Construction)

ItemCriteria
User needs documentedAll clinical needs captured with acceptance criteria
Requirements completeSRS covers all 9 categories (functional, performance, interface, cybersecurity, usability, risk control, data integrity, PHI, regulatory)
Risk management planISO 14971 plan approved, initial hazard analysis complete
Regulatory strategyPathway selected, predicates identified
Traceability (initial)UN → REQ links established

DR2 — Pre-Validation (After Construction, Before V&V)

ItemCriteria
Architecture documentedIEC 62304 §5.3 satisfied
SOUP assessedAll 3rd-party items risk-classified
Code reviews completeAll units reviewed, findings resolved
Design freezeNo more code changes during V&V
Risk controls implementedAll controls in hazard analysis coded + unit tested
SBOM generatedCycloneDX/SPDX with all dependencies

DR3 — Pre-Submission (After V&V)

ItemCriteria
All tests pass0 open failures on safety/efficacy tests
Coverage metClass-appropriate thresholds achieved
Residual risk acceptableBenefit-risk documented and favorable
Traceability closedFull bidirectional, no orphans
Anomalies dispositionedAll defects resolved or risk-accepted
Labeling reviewedIntended use, contraindications, IFU complete

11. FDA vs EU MDR Comparison for SaMD

AspectFDA (US)EU MDR 2017/745
ClassificationRisk-based (Class I/II/III)Rule 11: SaMD is IIa minimum; IIb/III if serious
Pathway510(k), De Novo, PMANotified Body conformity assessment
Clinical evidenceSubstantial equivalence (510k) or clinical studies (PMA)Clinical Evaluation Report ALWAYS required
Post-marketAnnual reports, MDRsPMCF + PSUR annually + vigilance
AI/ML updatesPCCP frameworkNotified Body re-assessment for significant changes
CybersecurityMandatory SBOM + threat model (2023)MDCG 2019-16 guidance (recommended)
QMS standard21 CFR 820 / QMSR (aligned with ISO 13485)ISO 13485 certification required
Software lifecycleIEC 62304 (recognized consensus standard)IEC 62304 (harmonized standard)
Timeline510(k): 3-6mo; De Novo: 9-12mo; PMA: 12-24moNB audit cycle: 12-18 months
Labeling21 CFR 801 + unique labeling requirementsAnnex I GSPR + SSCP (for implants/Class III)

Key gotcha: EU MDR Rule 11 classifies most SaMD as IIa minimum (vs FDA Class II). SaMD providing information for diagnosis of serious conditions = Class IIb or III in EU even if Class II in US. Plan for the higher classification upfront if dual-market.

12. Common SaMD Compliance Mistakes (Severity-Ranked)

  1. Wrong: Treating IEC 62304 safety class as equivalent to FDA device class Right: IEC 62304 Class A/B/C is SOFTWARE safety; FDA Class I/II/III is DEVICE risk. A Class II device can have Class C software. Why: Incorrect classification leads to insufficient documentation and testing — Critical severity

  2. Wrong: Starting code before design inputs are documented Right: Document user needs → derive requirements → get DR1 approval → THEN build Why: 21 CFR 820.30(c) requires design input documentation before design output. FDA will cite this. — Critical severity

  3. Wrong: Omitting SOUP from the risk analysis Right: Every SOUP item must be assessed for known anomalies that could contribute to hazards Why: IEC 62304 §7.1.3 mandates SOUP risk assessment; unassessed SOUP = undocumented risk — High severity

  4. Wrong: Using "testing" as the only risk control Right: Testing VERIFIES a risk control works — it is not itself a control. Controls are architectural (watchdogs, bounds checks, redundancy). Why: Confusing verification with mitigation leaves residual risk unaddressed — High severity

  5. Wrong: No traceability between risk controls and verification tests Right: Each risk control in the hazard analysis must link to a specific test proving it works Why: Without this link, you cannot demonstrate risk controls are effective — submission deficiency — High severity

  6. Wrong: Performing design reviews as rubber-stamp exercises Right: Design reviews must have documented attendees, specific findings, action items, and resolution evidence Why: FDA inspectors look for "objective evidence" of review — generic sign-offs get 483s — High severity

  7. Wrong: Assuming De Novo is slower/harder than 510(k) for novel devices Right: De Novo creates a new predicate, establishing your market category. Average review: 9-12 months. Why: Forcing substantial equivalence to a poor predicate leads to Additional Information requests and delays — Medium severity

  8. Wrong: Treating cybersecurity as optional for non-connected devices Right: FDA's 2023 guidance applies to ALL devices with software. Threat model scope includes supply chain, update mechanism, and data-at-rest. Why: FDA will RTA submissions without cybersecurity documentation — Medium severity

13. Part 11 (Electronic Records) Decision Tree

Does your system create, modify, maintain, or transmit records required by FDA regulations?
├─ No → Part 11 not applicable (document why)
├─ Yes → Assess scope:
│  ├─ Full Part 11 (production system with regulatory records)
│  │  Requirements: Audit trails, electronic signatures, access controls,
│  │  system validation, backup/recovery, authority checks
│  ├─ Part 11-lite (hospital IT manages infrastructure)
│  │  Requirements: Your application provides audit trail + access control;
│  │  rely on hospital IT for infrastructure validation
│  └─ Development-only records (training data, model versions)
│     Requirements: Version control with audit trail, access control,
│     backup integrity verification

Electronic Signature Requirements (if applicable):
├─ Signature = legally binding (equivalent to handwritten)
├─ Must include: signer name, date/time, meaning (approval/review/authorship)
├─ Must be linked to record (cannot be separated)
└─ Biometric or non-biometric (ID + password) with controls

14. Post-Market Surveillance Essentials

Mandatory Activities by Market

ActivityFDAEU MDRFrequency
Medical Device Report (MDR/vigilance)✅✅Within 30 days (FDA) / immediately for serious (EU)
Annual report✅Annually
Periodic Safety Update Report (PSUR)✅Annually (Class IIa+)
Post-Market Clinical Follow-up (PMCF)✅Ongoing
Complaint handling✅✅Per event
CAPA process✅✅Per finding
Trend analysis✅✅Quarterly minimum
Software update reporting✅✅Per update (risk-based)

CAPA Trigger Decision Tree

Signal detected (complaint, trend, field event)
├─ Safety-related?
│  ├─ Yes → Immediate CAPA + potential field corrective action
│  └─ No → Evaluate below
├─ Systematic (affects multiple units/users)?
│  ├─ Yes → CAPA (root cause likely systemic)
│  └─ No → Correction only (isolated incident)
├─ Recurring (≥3 occurrences same root cause)?
│  └─ Yes → CAPA mandatory (correction alone is insufficient)
└─ Regulatory requirement (audit finding, FDA warning letter)?
   └─ Yes → CAPA mandatory with defined timeline

When NOT to Use This Skill

  • General software engineering without medical device claims — design controls add overhead without regulatory value
  • Clinical data standards (CDISC SDTM/ADaM) — use cdisc-compliance skill instead
  • Wellness/fitness apps with NO diagnostic or therapeutic claims — not regulated as devices
  • eCTD submission assembly or eSTAR portal mechanics — this skill covers device compliance, not submission logistics
  • Manufacturing quality (GMP, process validation for physical devices) — this skill focuses on software lifecycle

When to Escalate to Human Expert

  • Safety classification disagreement where reasonable arguments support multiple classes — requires cross-functional risk team decision
  • Predicate selection for 510(k) where no clear substantial equivalence exists — may need pre-submission meeting with FDA
  • Clinical evidence sufficiency questions — requires clinical affairs and biostatistics expertise
  • EU MDR classification under Rule 11 for borderline CDS/wellness software — requires Notified Body pre-assessment
  • Post-market safety signal assessment for potential field action — requires medical director and regulatory affairs

FAQ

What does the medical-device-software-compliance skill do?

Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure, 510(k)/De Novo/PMA pathway selection, SOUP/OTS assessment, verification and validation planning, cybersecurity for medical devices, PCCP for AI/ML devices, design reviews, traceability matrices, or post-market surveillance. Triggers include "SaMD", "medical device software", "IEC 62304", "ISO 14971", "ISO 13485", "21 CFR 820", "QMSR", "510(k)", "De Novo", "PMA", "design history file", "DHF", "software safety classification", "SOUP list", "risk management", "hazard analysis", "design review", "V&V protocol", "FDA submission", "EU MDR", "clinical evaluation", "PCCP", "predetermined change control", "GMLP", "cybersecurity premarket", "Part 11", "design controls", "design inputs", "design outputs", "traceability matri

How do I install medical-device-software-compliance?

Run npx skills add https://github.com/awslabs/hcls-agent-skills --skill medical-device-software-compliance in your project. The skills CLI detects the agents you use and copies the skill into the right folder. For Claude Code you can also place the skill folder at ~/.claude/skills/medical-device-software-compliance/ (personal) or .claude/skills/medical-device-software-compliance/ (project).

Which AI agents can use medical-device-software-compliance?

Any agent that supports the open Agent Skills format (a folder with a SKILL.md file), including Claude Code, Codex, Cursor, GitHub Copilot, Gemini CLI, Windsurf, Cline, OpenCode and more.

Is medical-device-software-compliance safe to install?

No partner security audit is available yet. Skills can run commands and scripts, so read the SKILL.md and any bundled files before installing.

Who maintains medical-device-software-compliance?

It is published by awslabs in awslabs/hcls-agent-skills. It has 89 recorded installs.

More from awslabs/hcls-agent-skills

View all →
awslabs avatar

ml-researcher

awslabs/hcls-agent-skills

Reason about ML experiment design for healthcare and life sciences data. Use when the user asks to design an ML study, choose a model for clinical/biomedical data, set up cross-validation, pick evaluation metrics, audit fairness, plan a regulatory submission, or critique an ML pipeline on EHR, medical imaging, genomics, molecules, or clinical text. Triggers include "design an ML experiment", "which model for this clinical data", "how should I split", "nested CV", "class imbalance", "AUROC vs AUPRC", "calibration", "decision curve", "net benefit", "TRIPOD+AI", "PROBAST", "CLAIM", "FDA SaMD", "PCCP", "GMLP", "site generalization", "temporal leakage", "scaffold split", "foundation model evaluation", "subgroup fairness", "is this model ready for deployment".

AI & Agents92 installs
awslabs avatar

dicom-processing

awslabs/hcls-agent-skills

DICOM and NIfTI medical image processing pipeline. Triggers on DICOM, NIfTI, dcm2niix, de-identification, pydicom, DICOM header, conversion, anonymization, BIDS, DICOM tags, medical image format conversion, "DICOM to NIfTI", "burned-in PHI", "SeriesInstanceUID", "nibabel", "DICOM anonymization".

Marketing & SEO90 installs
awslabs avatar

clinical-data-standards

awslabs/hcls-agent-skills

Reason about clinical data terminology standards — MedDRA hierarchy (LLT→PT→HLT→HLGT→SOC), ICD-10 code structure and grouping, SNOMED CT concept model, LOINC panel relationships, and mapping decisions between systems. Use when the user asks to code adverse events, map diagnoses to ICD-10, choose a coding granularity level, group AEs by SOC or PT, interpret SNOMED CT relationships, select LOINC codes for lab panels, convert between terminology systems, or decide when to aggregate at HLT vs PT level. Triggers include "MedDRA coding", "ICD-10 grouping", "SNOMED CT", "LOINC panel", "adverse event coding", "terminology mapping", "SOC table", "preferred term", "code hierarchy", "clinical coding", "AE frequency table", "diagnosis grouping", "lab code selection", "cross-walk between terminologies".

AI & Agents90 installs
awslabs avatar

radiology-preprocessing

awslabs/hcls-agent-skills

Structural MRI/CT preprocessing pipeline for radiology workflows covering skull stripping, bias field correction, registration, and intensity normalization. Triggers on skull stripping, bias correction, registration, ANTs, FSL, HD-BET, N4, brain extraction, normalization, FLIRT, FNIRT, SyN, fslreorient2std, MNI registration, T1 preprocessing.

AI & Agents89 installs
awslabs avatar

edc-data-validation

awslabs/hcls-agent-skills

Generate code for EDC export validation, clinical data range checks, cross-form consistency checks, SDTM structure validation, controlled terminology verification, and define.xml generation. Use when the user asks to validate clinical trial data exports, check vital sign or lab value ranges, verify AE date consistency, validate SDTM datasets against CDISC rules, generate define.xml, or build an automated data review pipeline. Triggers include "EDC validation", "range check clinical data", "cross-form consistency", "SDTM validation", "controlled terminology check", "define.xml generation", "Medidata Rave export", "Oracle InForm", "Veeva Vault CDMS", "clinical data cleaning", "edit check", "data query", "SDTM structure check", "lab range check", "vital signs validation", "AE date check", "protocol deviation detection", "OpenCDISC", "Pinnacles 21", "P21 validation".

AI & Agents87 installs
awslabs avatar

ehr-data-parsing

awslabs/hcls-agent-skills

Parse and extract clinical data from HL7v2 messages and FHIR R4 resources using Python. Use when the user mentions HL7v2, HL7, FHIR, PID segment, OBX segment, MSH segment, Patient resource, Observation resource, Condition resource, MedicationRequest, EHR data extraction, clinical message parsing, FHIR bundle, HL7 to FHIR conversion, ADT message, ORU message, lab result extraction, or clinical data quality checks. Triggers include "parse HL7", "extract FHIR", "HL7v2 message", "FHIR resource", "PID segment", "OBX segment", "Patient resource", "Observation resource", "EHR parsing", "clinical data extraction", "HL7 to CSV", "FHIR to DataFrame".

Data & Databases87 installs

Popular education & science skills

View category →
mattpocock avatar

teach

mattpocock/skills

Teach the user a new skill or concept, within this workspace.

Education & Science973.7K installs
mattpocock avatar

scaffold-exercises

mattpocock/skills

Create exercise directory structures with sections, problems, solutions, and explainers that pass linting. Use when user wants to scaffold exercises, create exercise stubs, or set up a new course section.

Education & Science553.8K installs
lllllllama avatar

paper-context-resolver

lllllllama/rigorpilot-skills

Rigor Paper Context helper for README-first deep learning repo reproduction. Use only when the README and repository files leave a narrow reproduction-critical gap and the task is to resolve a specific paper detail such as dataset split, preprocessing, evaluation protocol, checkpoint mapping, or runtime assumption from primary paper sources while recording conflicts. Do not use for general paper summary, repo scanning, environment setup, command execution, title-only paper lookup, or replacing README guidance by default.

Education & Science503.2K installs
juliusbrussee avatar

caveman-learn

juliusbrussee/caveman

Act on a Caveman learn report - review the ranked token sinks, apply cost-lowering fixes with per-edit consent, and report what those fixes returned. Use when asked to lower an agent's token cost, what caveman has saved, to trim a heavy CLAUDE.md, or to offload re-pasted context into cavemem.

Education & Science132.9K installs
greensock avatar

gsap-plugins

greensock/gsap-skills

Official GSAP skill for GSAP plugins — registration, ScrollToPlugin, ScrollSmoother, Flip, Draggable, Inertia, Observer, SplitText, ScrambleText, SVG and physics plugins, CustomEase, EasePack, CustomWiggle, CustomBounce, GSDevTools. Use when the user asks about a GSAP plugin, scroll-to, flip animations, draggable, SVG drawing, or plugin registration.

Education & Science83.7K installs
pbakaus avatar

teach-impeccable

pbakaus/impeccable

DEPRECATED: Use /impeccable teach instead. This command has been folded into the impeccable skill.